Various errata for the HP5370B manual:
p 3-24, “TB1”
The manual states that this is only good for timeintervals < 320ps. This should read 320µs.
p 8-101, “LA0”, “LEN0”, “LEN1” & “LEN2”
These signals are mislabeled. Signal labeled LA0 is LEN0, LEN0 is LEN1, LEN1 is LEN2 and LEN2 is LA0. (See also: 8-99)